An evaluation of two commercial test procedures for the detection of lupus anticoagulant.
cross_sectional · Level IV
Where this comes from
- Record sourced from PubMed, PMID 7817935.
- No licence information is recorded for this record.
- Because redistribution is not established, this page shows the abstract only. Follow the links below for the full text.
Abstract
The authors have evaluated two commercial assays for lupus anticoagulants (LA), the DVVtest+DVV confirm (American Diagnostica, Greenwich, CT) and the PTT-LA+Staclot LA (Diagnostica Stago, Asnieres-Sur-Seine, France). The assays were compared with a previously described quantitative, semi-automated, and computer-assisted test, which is based on two assays, using the activated partial thromboplastin time (APTT) and the Russell's viper venom time (RVVT), respectively. Thirty well-characterized LA-positive and 30 LA-negative plasma samples were used to test the commercial assays. In some samples, both commercial tests failed to detect weak or moderately strong LAs. The sensitivity was 80% for the DVVtest+DVV confirm and 67% for the PTT-LA+Staclot LA. However, by applying both the two commercial tests to all plasma samples, 29 of 30 LA-positive plasma samples were diagnosed (sensitivity 97%). No false positive test results were found (specificity 100%).
Medical subject headings
- Blood Coagulation Tests
- Lupus Coagulation Inhibitor