Clinical evaluation of pulmonary nodules with single-exposure dual-energy subtraction chest radiography with an iterative noise-reduction algorithm.
cross_sectional · Level IV
Where this comes from
- Record sourced from PubMed, PMID 7824718.
- No licence information is recorded for this record.
- Because redistribution is not established, this page shows the abstract only. Follow the links below for the full text.
Abstract
To compare the clinical usefulness of the single-exposure dual-energy subtraction method with an iterative noise-reduction algorithm. Fourteen radiologists read three sets of images from 44 patients: original computed radiographic images only, original computed radiographic images plus conventional bone-subtracted images, and original computed radiographic images plus iterative noise-reduced bone-subtracted images. Twenty-two patients had one or more (maximum, five) pulmonary nodules; 22 had no pulmonary nodules. Observer performance was evaluated by means of calculation of the average area under the alternative free-response receiver operating characteristic curves (A1). Compared with the original computed radiographic image only, detection of nodules was significantly better with both the original computed radiographic image plus iterative bone-subtracted image (A1 = 0.72 +/- 0.02 and 0.66 +/- 0.02, respectively; P = .01) and the original computed radiographic image plus conventional bone-subtracted image (A1 = 0.66 +/- 0.02 and 0.61 +/- 0.01, respectively; P = .03). The iterative noise-reduction algorithm is superior to conventional methods in detection of pulmonary nodules.
Medical subject headings
- Lung Diseases
- Radiography, Dual-Energy Scanned Projection
- Radiography, Thoracic
- Subtraction Technique