A universal nitric oxide delivery system.
basic_science · Level V
Where this comes from
- Record sourced from PubMed, PMID 7826805.
- No licence information is recorded for this record.
- Because redistribution is not established, this page shows the abstract only. Follow the links below for the full text.
Abstract
A nitric oxide delivery system for adult ventilators has been developed; this is designed to maintain a constant inspired nitric oxide concentration of 16 volumes per million (vpm) in spite of changes in minute and tidal volumes. In a laboratory test the system maintained tracer gas concentrations equivalent to nitric oxide 13.8-18 vpm for tidal volumes of 500, 700 and 900 ml, ventilator rates of 10, 15 and 20 b.p.m., peak inspiratory flow rates of 30, 40 and 50 litre min-1 and square, sine and decelerating ramp flow profiles.
Medical subject headings
- Anesthesia, Inhalation
- Nitric Oxide