Multiplane transesophageal echocardiographic doppler imaging accurately determines cardiac output measurements in critically ill patients.

Feinberg, M S; Hopkins, W E; Davila-Roman, V G; Barzilai, B · Chest · 1995

prospective_cohort · Level II

Where this comes from

Abstract

OBJECTIVES: To compare cardiac output and stroke volume measured by multiplane transesophageal Doppler echocardiography with that measured by the thermodilution technique. DESIGN: Prospective direct comparison of paired measurements by both techniques in each patient. SETTING: Cardiac surgery and myocardial infarction intensive care units. PATIENTS: Twenty-nine patients, mean age (+/- SD) 67 +/- 8 years. Nineteen had undergone open heart surgery and 10 had suffered acute myocardial infarction. METHODS: Cardiac output and stroke volume were measured simultaneously by the thermodilution technique and multiplane transesophageal Doppler echocardiography via the transgastric view (119 +/- 8 degrees) with the sample volume positioned at the level of the left ventricular outflow tract. RESULTS: Stroke volume and cardiac output measurements were obtained in 29 of 33 patients (88%). Mean values were 50 +/- 13 mL and 4.8 +/- 1.3 L/min by Doppler and 51 +/- 14 mL and 4.9 +/- 1.4 L/min by thermodilution (r = 0.90, r = 0.91, p < 0.001). The mean differences in values obtained with the two techniques were 1 +/- 6 mL (2 +/- 12%) and 0.1 +/- 0.7 L/min (2 +/- 12%). CONCLUSIONS: Multiplane transesophageal echocardiography enhances the ability to estimate accurately cardiac output and stroke volume by providing new access to left ventricular outflow tract in critically ill patients.