Lateral stability of proximal interphalangeal joint replacement.
biomechanical · Level V
Where this comes from
- Record sourced from PubMed, PMID 7876481.
- No licence information is recorded for this record.
- Because redistribution is not established, this page shows the abstract only. Follow the links below for the full text.
Abstract
The lateral stability of the cadaver proximal interphalangeal joint was studied with an electromagnetic tracking system before and after implant replacement. Ten middle fingers were retrieved and randomly divided into two groups for joint replacement with either a silicone rubber flexible finger joint implant or a custom surface replacement prosthesis. Each finger was mounted on a fixture that allowed loading of the proximal interphalangeal joint with tension through the flexor, extensor, and intrinsic tendons. For the intact proximal interphalangeal joint under lateral stress, lateral angulation averaged 4 degrees in extension and 8 degrees with the joint in 60 degrees of flexion. Although lateral angulation increased after surface replacement prosthesis implantation when the joint was flexed more than 20 degrees, comparison with the intact joint showed no statistical difference. With the silicone implant, lateral angulation became more pronounced even in the extended position and showed a significant difference when the proximal interphalangeal joint was flexed more than 20 degrees.
Medical subject headings
- Joint Prosthesis
- Metacarpophalangeal Joint
- Silicone Elastomers
Anatomy
- hand