Patient positioning for fractionated precision radiation treatment of targets in the head using fiducial markers.
other
Where this comes from
- Record sourced from PubMed, PMID 7878212.
- No licence information is recorded for this record.
- Because redistribution is not established, this page shows the abstract only. Follow the links below for the full text.
Abstract
When irradiating targets in the brain, an accurately localised dose is often needed. One crucial moment to achieve this is the positioning of the patient. We have developed a positioning method where the patient is immobilised with a bite block and a head mould, and where the position of the target is determined by X-ray imaging of fiducial markers that are placed in the patient's skull. A method for computing the transformation needed to move the target from the observed to the prescribed position and orientation is described. This method uses the information from two orthogonal X-ray images and takes measurement errors and data from three or more markers into account. Results from using the method clinically in proton beam therapy are given.