Right atrial compression related to defibrillator patches.
case_report · Level V
Where this comes from
- Record sourced from PubMed, PMID 7887712.
- No licence information is recorded for this record.
- Because redistribution is not established, this page shows the abstract only. Follow the links below for the full text.
Abstract
Acceptable function of an internal defibrillator can be achieved with different patch orientations. For patients requiring defibrillator patches concomitant with a cardiac procedure requiring extracorporeal circulation, application of one of the patches within the pericardium adjacent to the right atrium has provided excellent defibrillation thresholds. We describe 4 such patients in whom a compressing thrombus subsequently developed between the patch and the atrium. The thrombus was small and asymptomatic in 1 patient, but caused localized tamponade requiring reexploration in 2 patients and a fatal superior vena caval obstruction in 1. The precise etiology of this serious complication remains unclear, but its occurrence argues against the application of intrapericardial defibrillator patches in this orientation.