A controller for automatic regulation of induced paralysis during surgery.
basic_science · Level V
Where this comes from
- Record sourced from PubMed, PMID 7915504.
- No licence information is recorded for this record.
- Because redistribution is not established, this page shows the abstract only. Follow the links below for the full text.
Abstract
This paper presents a strategy for automatic control of induced paralysis using vecuronium bromide during surgery. The controller is self-tuning and adapts to inter-patient and intra-patient response variations while optimizing the output variance and infusion rate. In particular, the controller is capable of accommodating the variations of pure time delays in patient response. The performance of the controller is evaluated using an experimentally derived pharmacokinetic and nonlinear pharmacodynamic model of patient response. The results indicate that the controller provides robust regulation of the paralysis level with no output offset.
Medical subject headings
- Algorithms
- Decision Making, Computer-Assisted
- Monitoring, Physiologic
- Paralysis
- Vecuronium Bromide