Temporal electrical heterogeneity for detecting coronary artery disease: results in a heterogeneous cardiac population.
prospective_cohort · Level II
Where this comes from
- Record sourced from PubMed, PMID 7923749.
- No licence information is recorded for this record.
- Because redistribution is not established, this page shows the abstract only. Follow the links below for the full text.
Abstract
BACKGROUND: A new 22-lead electrocardiographic test has been advocated as a screening tool for coronary artery disease and has been shown to have accuracy similar to stress electrocardiography in specific patient populations. OBJECTIVE: To determine the accuracy of this test for detecting coronary artery disease in patients undergoing coronary angiography for a variety of cardiac conditions. METHODS: We prospectively determined the temporal electrical heterogeneity (TEH) index at rest in 70 patients who had no angina or Q waves on the resting 12-lead electrocardiogram before they underwent coronary angiography. RESULTS: Twenty-six of the 70 patients had significant coronary artery disease, defined as 70% stenosis or greater in at least one major epicardial coronary artery. A TEH index of 80 or more had a sensitivity of 58%, a specificity of 75%, and a positive predictive value of 58%. The group with significant coronary disease had a mean TEH index of 77.2, and the group without coronary disease had a mean index of 65.5 (P = .02), despite similar clinical characteristics and indications for angiography. CONCLUSION: The TEH index shows promise as a screening tool for coronary artery disease in a heterogeneous cardiac population. However, larger studies are needed before it can be endorsed for widespread clinical use.