ACT3: a high-speed, high-precision electrical impedance tomograph.
basic_science · Level V
Where this comes from
- Record sourced from PubMed, PMID 7927393.
- Also identified by PMC identifier 4793976.
- No licence information is recorded for this record.
- Because redistribution is not established, this page shows the abstract only. Follow the links below for the full text.
Abstract
This paper presents the design, implementation, and performance of Rensselaer's third-generation Adaptive Current Tomograph, ACT3. This system uses 32 current sources and 32 phase-sensitive voltmeters to make a 32-electrode system that is capable of applying arbitrary spatial patterns of current. The instrumentation provides 16 b precision on both the current values and the real and reactive voltage readings and can collect the data for a single image in 133 ms. Additionally, the instrument is able to automatically calibrate its voltmeters and current sources and adjust the current source output impedance under computer control. The major system components are discussed in detail and performance results are given. Images obtained using stationary agar targets and a moving pendulum in a phantom as well as in vivo resistivity profiles showing human respiration are shown.
Medical subject headings
- Signal Processing, Computer-Assisted
- Tomography