Biomechanical factors in designing screws for the Orthofix system.
biomechanical · Level V
Where this comes from
- Record sourced from PubMed, PMID 7955702.
- No licence information is recorded for this record.
- Because redistribution is not established, this page shows the abstract only. Follow the links below for the full text.
Abstract
In a monolateral external fixation system, bone grip is achieved optimally using half pins. This article reports the mathematical modeling used to design the screws and the drills of the Orthofix system. The diameter of the shank must be such that there is no risk of fracture at the screw entry point. It must also provide high strength under weight-bearing deformation stresses. The shape of the threaded part must allow the highest possible bone-gripping strength while minimizing slackening in relation to external and internal loads. The thermal effects of drilling and screw insertion must be considered to minimize necrotic damage that may stimulate reabsorption phenomena around the screw. The design of the different components of the pins must vary according to all the above factors to ensure optimal interaction between the fixator, screw, and bone.
Medical subject headings
- Bone Screws
- External Fixators