Are pulse oximetry and end-tidal carbon dioxide tension monitoring reliable during laparoscopic surgery?
case_series · Level IV
Where this comes from
- Record sourced from PubMed, PMID 7978132.
- No licence information is recorded for this record.
- Because redistribution is not established, this page shows the abstract only. Follow the links below for the full text.
Abstract
Cardiorespiratory changes induced by pneumoperitoneum and head-up tilt may generate alveolar ventilation to perfusion ratio changes and increased systemic vascular resistances. The reliability of end-tidal carbon dioxide tension and pulse oximetry in predicting arterial carbon dioxide partial pressure and arterial oxygen saturation may therefore be affected. The 35 ASA 1-2 patients in this study comprised 12 men and 23 women aged 48 (SD 17) years and weighing 71 (SD 14) kg. Twenty-nine were to undergo upper abdominal laparoscopy for cholecystectomy and six hyperselective vagotomy. Intra-abdominal pressure was 1.7 (SD 0.9) kPa and head-up tilt was 5.6 (SD 4.2) degrees. After abdominal insuflation, arterial carbon dioxide partial pressure significantly increased (p < 0.05). However, the arterial carbon dioxide partial pressure-end-tidal carbon dioxide partial pressure gradient remained constant throughout surgery. This gradient was highly correlated with arterial carbon dioxide partial pressure (p < 0.0001), but was not correlated with elapsed time, intra-abdominal pressure or head-up tilt. Arterial oxygen saturation was always greater than 95% in all patients and the arterial oxygen saturation-pulse oximetric saturation gradient was always less than or equal to +4%. In conclusion, end-tidal carbon dioxide partial pressure and pulse oximetric saturation allow reliable monitoring of arterial carbon dioxide partial pressure and arterial oxygen saturation in the absence of pre-existing cardiopulmonary disease and/or acute peroperative disturbance.
Medical subject headings
- Carbon Dioxide
- Laparoscopy
- Oximetry
- Tidal Volume