The effect of trochlear design on patellofemoral shear and compressive forces in total knee arthroplasty.
biomechanical · Level V
Where this comes from
- Record sourced from PubMed, PMID 7994950.
- No licence information is recorded for this record.
- Because redistribution is not established, this page shows the abstract only. Follow the links below for the full text.
Abstract
Biomechanical testing was performed on 5 cadaveric knees to evaluate the effect of patellofemoral design on shear and compressive force at the patellar component-bone interface. Three patellofemoral designs with identical tibiofemoral articular surfaces were tested. The knees were tested under dynamic loading conditions from 0 degree to 100 degrees flexion. A transducer that measured 3 orthogonal force components was mounted between the patellar component and the patella. The combination of an oval dome patella with a 2-mm deeper trochlear groove and associated increased femoral resection was compared with a biconcave patella with standard trochlea and normal femoral resection. Alignment of the trochlear groove in 7 degrees valgus (anatomic) decreased medial-lateral shear by 10% to symmetric trochlear groove alignment.
Medical subject headings
- Knee Prosthesis
- Prosthesis Design
- Stress, Mechanical
Anatomy
- femur
- knee