Pulse oximetry is accurate in acute anemia from hemorrhage.
case_series · Level IV
Where this comes from
- Record sourced from PubMed, PMID 8010546.
- No licence information is recorded for this record.
- Because redistribution is not established, this page shows the abstract only. Follow the links below for the full text.
Abstract
Pulse oximetry is an accurate, noninvasive assay of oxygen saturation percentage (SaO2) in acutely severe anemia (less than 5 g/dL). A paired comparison of SaO2 by pulse oximetry (SpO2) and arterial blood gas analysis (SaO2) in patients with either acute gastrointestinal bleeding or blunt trauma with hematocrit of less than 20%. An academic emergency department in a tertiary referral hospital with a Level I trauma center. Seventeen nonhypoxic patients were identified with initial hematocrit of less than 20% over a 9-month period. Determination of SaO2 by pulse oximetry and arterial blood analysis was not statistically significantly different (P < .05) over a range of hemoglobin levels from 2.3 to 8.7 g/dL. The observed mean difference between SpO2 and SaO2 was 0.53 +/- 0.23% (mean +/- SEM; n = 17). The 95% confidence interval associated with this difference score ranged from 0.044 to 1.014. Despite reports that pulse oximetry is inaccurate below a hemoglobin concentration of 5 g/dL, our study suggested that this technology is accurate and reliable at 2.3 g/dL for nonhypoxic SaO2 values (SaO2 of more than 93%). The anemic endpoint where pulse oximetry either becomes inaccurate or simply fails to work has not been determined. Pulse oximetry overestimates SaO2 on average by 0.53% over a range of hemoglobin concentrations from 2.3 to 8.7 g/dL.
Medical subject headings
- Anemia
- Hemorrhage
- Oximetry
- Oxygen