An evaluation of the microbial retention performance of three ventilator-circuit filters.
rct · Level II
Where this comes from
- Record sourced from PubMed, PMID 8014294.
- No licence information is recorded for this record.
- Because redistribution is not established, this page shows the abstract only. Follow the links below for the full text.
Abstract
to determine and compare the microbial retention efficiency of 3 types of heat and moisture exchange (HME) filters. randomised prospective study. Intensive Care Unit at the Middlesex Hospital and the microbiology laboratory of the Department of Academic Microbiology, University College London Medical School. An aerosol challenge to each filter using a suspension containing. S. marcescens and the bacteriophage MS2. In total 45 filters of 3 types were tested, 15 previously unused and 30 that had been used in ventilator circuits of 30 patients. The 3 types of filter generally had microbial retention efficiencies of > 99.99% for both S. marcescens and the bacteriophage MS2. The 3 types of HME filter were effective microbial filters and comparable in their microbial retention.
Medical subject headings
- Equipment Contamination
- Filtration
- Levivirus
- Respiration, Artificial
- Serratia marcescens