Defibrillator twiddler's syndrome.
case_report · Level V
Where this comes from
- Record sourced from PubMed, PMID 8037540.
- No licence information is recorded for this record.
- Because redistribution is not established, this page shows the abstract only. Follow the links below for the full text.
Abstract
Long-term complications of internal cardioverter defibrillators have generally involved inappropriate shocks or hardware malfunctions. The present case documents a new problem of internal cardioverter defibrillator lead disruption resulting from patient-induced rotation of the generator in its pocket, similar to the previously reported pacemaker twiddler's syndrome. However, unlike the usual early symptoms of pacemaker nonfunction, the first symptomatic manifestation of the defibrillator twiddler's syndrome may be an inappropriate internal cardioverter defibrillator discharge or more importantly, unrescued sudden death. Careful attention to patient complaints about the internal cardioverter defibrillator generator and being alert to subtle lead changes on chest radiographs may allow the clinician to recognize this syndrome before a major, possibly fatal event.
Medical subject headings
- Defibrillators, Implantable