Technique for mechanical measurements using optical scattering from a micropipette.
biomechanical · Level V
Where this comes from
- Record sourced from PubMed, PMID 8045585.
- No licence information is recorded for this record.
- Because redistribution is not established, this page shows the abstract only. Follow the links below for the full text.
Abstract
This communication proposes an optical method using forward light scattering to measure the deflection of a micropipette from a normal contact force with a small elastic solid, and the damping coefficient of a small fluid sample through a dynamic interaction with the pipette. An analytical and experimental study of this sensor is performed and described for each property. Topics covered include: 1) forward light scattering technique for determining the diameter of a micropipette (using exact and simplified geometric interference theories); 2) deflection measurement and analysis on the pipette-sample system using the finite element method; and 3) equivalent damping coefficient and resonant frequency measurement of the pipette-fluid system and its numerical solution derived from the finite element model. A comparison of the experimental results and the analytical results is included to provide the means for analyzing the potential of this new method.
Medical subject headings
- Micromanipulation
- Optics and Photonics
- Robotics