Discrepancies in anal manometric pressure measurement--important or inconsequential?

Morgado, P J; Wexner, S D; Jorge, J M · Dis Colon Rectum · 1994

cross_sectional · Level IV

Where this comes from

Abstract

Maximum resting and squeeze pressures have been the most widely employed parameters for manometric assessment of the anal sphincters. However, a single maximum value may not always be the best assessment. The aim of this study was to compare mean and maximum resting and mean and maximum squeeze pressures in a large sample population. All manometric pressure profiles were reviewed by a single individual blinded to the patient's age and diagnosis. Four hundred sixty-six patients with a measurable high-pressure zone were included in this study. The study population was comprised of 279 females and 186 males. A significant difference was found between mean (56.26 mmHg) and maximum (79.2 mmHg) resting pressures (P < 0.01) and also between mean (81.25 mmHg) and maximum (119.50 mmHg) squeeze pressures (P < 0.01). A significant difference (P < 0.01) was also observed when compared by length of the high-pressure zone. The measurement, documentation, and reporting of mean resting and mean squeeze pressures provide a better perspective of anal manometric results, since the two sets of values are significantly different (P < 0.01), regardless of the anal canal length. Therefore, these data support the standardized evaluation of both mean and maximum pressures in individual patients and in published series.

Medical subject headings