Alternative method of ultrafiltration after cardiopulmonary bypass.

Groom, R C; Akl, B F; Albus, R A; Hill, A; Munoz, R; Lefrak, E A · Ann Thorac Surg · 1994

case_series · Level IV

Where this comes from

Abstract

A revised circuit design for modified ultrafiltration is presented rendering the technique more convenient for use after cardiopulmonary bypass when blood cardioplegia is used. The procedure employs a hollow-fiber ultrafiltration device attached to the cardioplegia circuit. A bubble trap, heat exchanger, and a pressure monitor are incorporated as safety features. The technique has been used in 80 patients (30 pediatric and 50 adult) and has been associated with relevant increases in colloid osmotic pressure and hematocrit.