Air cavity effects on the radiation dose to the larynx using Co-60, 6 MV, and 10 MV photon beams.
biomechanical · Level V
Where this comes from
- Record sourced from PubMed, PMID 8083084.
- No licence information is recorded for this record.
- Because redistribution is not established, this page shows the abstract only. Follow the links below for the full text.
Abstract
To determine the perturbation effect in the surface layers of lesions located in the air-tumor tissue interface of larynx using 60Co, 6 MV, and 10 MV photon beams. Thermoluminescent dosimeters (TLDs), were embedded at 16 measurement locations in slab no. 8 of a humanoid phantom and exposed to two lateral-opposed beams using standard small 7 x 7 cm fields. Similarly, radiographic and radiochromic films were placed between slabs no. 7 and no. 8 of the humanoid phantom and exposed to two lateral-opposed radiation beams. The dosimeters were irradiated with 60Co, 6 MV, and 10 MV photon beams. Computer tomography (CT) treatment planning without inhomogeneity correction was performed. At the tissue-air interface, the average measured percentage dose (% dosem) is about (108.7 +/- 4.8)% with TLD data, (96.8 +/- 2.5)% with radiographic film data, and (100.8 +/- 4.9)% with radiochromic film data. Similarly, in the central part of the cavity, the % dosem is (98.4 +/- 3.1)% with TLD data, (94.3 +/- 3.3)% with radiographic film data, and (91.7 +/- 5.0)% with radiochromic film data. Using the CT-based generated dose distribution (without inhomogeneity correction), the average calculated percentage dose (% dosec) is (98.7 +/- 1.0)% at the tissue-air interface and 98% in the central part of the air cavity. For the beam energies studied, the variation from the % dosem at the tissue-air interface for a given dosimetry technique is relatively small [< 5% (TLD), < 3% (radiographic), and < 5% (radiochromic)] and therefore should not be significant in clinical settings. The variation from the % dosem at the tissue-air interface is more significant for lower energies [8% (60Co), 7.3% (6 MV)]. This variation is about 4.3% for 10 MV photon beam, therefore, while our institutional practice favors lower energy (60Co to 6 MV) for node-negative glottic cancers, physical/dosimetric evidence offers no disadvantage to the use of higher energy photons.
Medical subject headings
- Larynx
- Radiotherapy Planning, Computer-Assisted
- Respiratory System