Imaging of cerebral activation at 1.5 T: optimizing a technique for conventional hardware.
basic_science · Level V
Where this comes from
- Record sourced from PubMed, PMID 8115643.
- No licence information is recorded for this record.
- Because redistribution is not established, this page shows the abstract only. Follow the links below for the full text.
Abstract
To empirically optimize a two-dimensional magnetic resonance (MR) imaging technique for detecting changes in signal intensity during cerebral activation with a standard clinical imager. Visual activation experiments were performed while imaging parameters were manipulated in a serial fashion, to test their effect on the percentage change in signal intensity (PCSI). The parameters tested were section thickness, echo time (TE), field of view (FOV), flip angle, radio-frequency (RF) spoiling, number of readout points, and number of signals averaged. The PCSI for visual activation experiments was typically in the 1%-5% range. The best results were achieved with a small section thickness, long TE, and large FOV. Variations in other acquisition parameters had a negligible influence on the PCSI. Good results can be obtained at functional MR imaging with standard 1.5-T hardware. Markedly improved results can be obtained by optimizing several key variables, namely, section thickness, TE, and FOV.
Medical subject headings
- Brain
- Cerebrovascular Circulation
- Magnetic Resonance Imaging