Ingrowth surfaces. Plasma spray coating to titanium alloy hip replacements.
review · Level V
Where this comes from
- Record sourced from PubMed, PMID 8118992.
- No licence information is recorded for this record.
- Because redistribution is not established, this page shows the abstract only. Follow the links below for the full text.
Abstract
Plasma spray coatings on titanium alloy total hip replacements (THRs) provide a viable alternative to sintered bead or diffusion-bonded fiber metal ingrowth surfaces. The plasma spray process enables the titanium alloy implant to retain 90% of its fatigue strength, as opposed to less than 50% for sintered or diffusion-bonded cementless devices. Laboratory and clinical evidence suggest that plasma spray cementless THRs perform as well, if not better, than other cementless implants. Clinical trials have demonstrated the superiority of a plasma spray cementless THR (Mallory-Head) over a sintered bead cementless THR (PCA) and the equivalent results to a cemented THR. Plasma spray coatings on titanium alloy hip replacements provide a safe, predictable coating for cementless fixation of THRs in the short to medium range of follow-up periods.
Medical subject headings
- Hip Prosthesis
- Osseointegration
- Titanium
Anatomy
- hip