Initial mechanical stability of acetabular prostheses.
biomechanical · Level V
Where this comes from
- Record sourced from PubMed, PMID 8121834.
- No licence information is recorded for this record.
- Because redistribution is not established, this page shows the abstract only. Follow the links below for the full text.
Abstract
Secure initial stability is crucial for prostheses designed for biological ingrowth; without it, subsequent tissue ingrowth is impossible. Mechanical testing was performed in vitro to measure the interface shear stability of acetabular cup prostheses. Cemented, threaded, fined, and press-fit cups, both with and without supplemental screw fixation, were tested. A mechanically consistent polyethylene foam was used as a substrate to obviate the variations inherent in cadaver bones. Ramped, eccentric rim loads were applied using a servohydraulic materials testing system. Reaming, insertion, and loading variables were carefully controlled. Cups with external threads were significantly more stable than those fixed by fins or screws. Press-fit cups were significantly looser than all other tested designs. None of the uncemented prostheses approached the initial stability of cups secured with acrylic bone cement.
Medical subject headings
- Acetabulum
- Prostheses and Implants
Anatomy
- hip
- pelvis