Comments on distinguishability in electrical impedance imaging.
Where this comes from
- Record sourced from PubMed, PMID 8125509.
- No licence information is recorded for this record.
- Because redistribution is not established, this page shows the abstract only. Follow the links below for the full text.
Abstract
In Electrical Impedance Imaging (EII), the spatial distribution of electrical resistivity of a volume conductor is reconstructed from measurements of potential which result from an externally applied electric field. The ability of such a system to distinguish a target inhomogeneity and the discussion of several measures of distinguishability has been a subject of a recent study. In this communication a few comments are added to those of. The medical device safety regulations limit the maximum total input current which can be applied to the human thorax. Based on the currently existing safety regulations, if the input current to the thorax is limited and constant, patterns applied between opposite electrodes result in higher distinguishability of a centered target than the cosine current patterns as suggested by.
Medical subject headings
- Electric Impedance
- Plethysmography, Impedance