Screening for insecticide overexposure under field conditions: a reevaluation of the tintometric cholinesterase kit.
case_control · Level III
Where this comes from
- Record sourced from PubMed, PMID 8129072.
- Also identified by PMC identifier 1614842.
- No licence information is recorded for this record.
- Because redistribution is not established, this page shows the abstract only. Follow the links below for the full text.
Abstract
A semiquantitative tintometric field kit has been used in the developing world for almost 30 years to measure whole blood cholinesterase levels in persons exposed to organophosphate pesticides. The validity of this screening kit was evaluated among 79 workers heavily exposed to organophosphates by comparison with a reference assay for erythrocyte cholinesterase. Overall correlation between the two methods was good. However, either sensitivity or specificity of the tintometric kit was less than 75% for each of the three tintometric categories commonly used to define the limit of normal. Because baseline erythrocyte cholinesterase levels were not available for this population, the true sensitivity and specificity of the tintometric assay may be even lower.
Medical subject headings
- Agriculture
- Cholinesterases
- Environmental Monitoring
- Insecticides
- Occupational Exposure
- Organophosphorus Compounds