The value of contrast arthrography in assessing loosening of symptomatic uncemented total hip components.
case_series · Level IV
Where this comes from
- Record sourced from PubMed, PMID 8160035.
- No licence information is recorded for this record.
- Because redistribution is not established, this page shows the abstract only. Follow the links below for the full text.
Abstract
Twenty-five patients with symptomatic uncemented total hip components were studied with contrast arthrography prior to surgical exploration. All but one had uncemented femoral stems and 16 had an uncemented acetabular component. As judged by the findings at surgery, on the femoral side the sensitivity, specificity, and accuracy of arthrography were 57%, 60%, and 58% respectively. There was a relatively high incidence of both false positives (17%) and false negatives (25%). On the acetabular side sensitivity, specificity, and accuracy were 29%, 89%, and 62.5%. False negatives were common (31%), while there was only one false positive. The results in this small series show that arthrography has distinct limitations in identifying the fixation status of uncemented total hip components.
Medical subject headings
- Contrast Media
- Hip Joint
- Hip Prosthesis
Anatomy
- hip