Measurement of dead-space in a model lung using an oscillating inspired argon signal.
basic_science · Level V
Where this comes from
- Record sourced from PubMed, PMID 8171946.
- No licence information is recorded for this record.
- Because redistribution is not established, this page shows the abstract only. Follow the links below for the full text.
Abstract
In a model lung, airways dead-space can be accurately measured using a forced inspired oscillating argon signal, which varies sinusoidally about a mean concentration of 6% v/v with an amplitude of +/- 4% v/v. With sinusoid forcing periods longer than 120 seconds, and at a breathing rate of 13.4 breaths minute-1, the mean airways dead-space can be measured with a standard error of less than 5%. Sinusoid forcing periods shorter than 120 s provided inaccurate estimates of dead-space and so should not be used with this technique.
Medical subject headings
- Argon
- Lung
- Models, Biological
- Respiratory Dead Space