Evaluation of neutron irradiation field for boron neutron capture therapy by using absorbed dose in a phantom.
basic_science · Level V
Where this comes from
- Record sourced from PubMed, PMID 8175399.
- No licence information is recorded for this record.
- Because redistribution is not established, this page shows the abstract only. Follow the links below for the full text.
Abstract
The method of an evaluation for the dose characteristics of BNCT is presented, and an eccentric core design for the TRIGA-II reactor is proposed. We have defined the "irradiation time" as the time of irradiation in which the "maximum 1 microgram/g dose" becomes 3,000 RBE-cGy, because we assumed that the normal tissue contained 1 microgram/g 10B. We have also changed the RBE values and calculated the absorbed dose in the "irradiation time" by using an arrangement including both a facility structure and a body phantom. Moreover, we have modified the dose criteria for BNCT as follows: The "eye dose," "total body dose," and "except-head dose" should be less than 200, 100, and 50 RBE-cGy, respectively. We have added one more criteria for BNCT-that the thermal neutron fluence at the tumor position (5 cm from the surface) should be over 2.5 x 10(12) n/cm2 in the "irradiation time." The distance from the core side to the irradiation port is a very important factor to design a neutron irradiation field for BNCT. We can get the acceptable dose for BNCT with only 1-h irradiation by using a 100 kW reactor if we can get the irradiation port at the distance of 120 cm from the core side.
Medical subject headings
- Boron Neutron Capture Therapy
- Neutrons
- Radiotherapy Dosage