Extracranial carotid artery flow measurement during carotid endarterectomy using a Doppler ultrasonographic flowmeter.

Whyman, M R; Naylor, A R; Ruckley, C V; Wildsmith, J A · Br J Surg · 1994

case_series · Level IV

Where this comes from

Abstract

A study was performed to evaluate immediate changes in carotid artery blood flow after carotid endarterectomy using a Doppler ultrasonographic flowmeter. Forty-seven consecutive patients aged 49-78 (median 64) years with symptomatic internal carotid artery (ICA) stenosis underwent carotid endarterectomy. Volume flow, resistance and pulsatility index were measured in the common carotid artery (CCA) and ICA both before and after operation. ICA flow increased by 54 per cent (P < 0.01) and resistance decreased by 58 per cent (P = 0.01). CCA flow rose by 17 per cent (P = 0.1) and resistance fell by 21 per cent (P = 0.02). The pulsatility index did not change significantly. The Doppler ultrasonographic flowmeter offers a simple, non-invasive and convenient means of improving understanding of the immediate haemodynamic effects of carotid endarterectomy. It is useful in everyday practice to confirm the adequacy of endarterectomy, particularly in the absence of other methods of quality control. Duplex ultrasonography might still be necessary after surgery if the flowmeter does not demonstrate an increase in flow compared with the preoperative value.