Reproducibility of postural changes in intraocular pressure with the Tono-Pen and Pulsair tonometers.
cross_sectional · Level IV
Where this comes from
- Record sourced from PubMed, PMID 8213979.
- No licence information is recorded for this record.
- Because redistribution is not established, this page shows the abstract only. Follow the links below for the full text.
Abstract
Reproducibility of measurements of postural changes in intraocular pressure was determined by using the Tono-Pen (Mentor O & O, Inc., Norwell, Massachusetts) and Pulsair (Keeler Instruments, Inc., Broomhall, Pennsylvania) tonometers. Thirty subjects had three repeated sitting and reclining measurements performed on three separate visit days. Reproducibility coefficients of 3.3% for the Tono-Pen and 6.3% for the Pulsair were obtained for the within-visit postural change measurements. Across-visit reproducibility coefficients were 7.9% and 26.2% for the Tono-Pen and Pulsair, respectively. The estimated standard deviations for both the within-visit and across-visit postural change measurements were high for both instruments. These results indicate poor reproducibility of measurements of postural changes in intraocular pressure.
Medical subject headings
- Intraocular Pressure
- Posture
- Tonometry, Ocular