Effect of motion on thallium-201 SPECT studies: a simulation and clinical study.
other
Where this comes from
- Record sourced from PubMed, PMID 8229222.
- No licence information is recorded for this record.
- Because redistribution is not established, this page shows the abstract only. Follow the links below for the full text.
Abstract
Although patient motion on 201Tl SPECT studies has been reported as a source of artifacts, systematic studies on motion patterns and resultant artifacts are lacking. Accordingly, we simulated 74 motion patterns upon a normal study. The tomograms were assessed for presence of defects: The "motion pixel area index" ranged from 1 to 83; 26 of 30 (87%) simulations with an index > or = 21 had defects, whereas 38 of 44 (86%) simulations with an index < 21 were normal. Defect location was dependent on motion direction; defect intensity was dependent on its magnitude and timing. Review of data acquisition in 164 recent normal patient studies revealed motion in 42 (26%). Motion was generally minimal and caused defects in only seven (4%). Thus, mild motion is unlikely to produce defects. In our laboratory, motion is now an infrequent source of artifacts; severe motion produces recognizable patterns that depend on its direction, magnitude and timing.
Medical subject headings
- Motion
- Thallium Radioisotopes
- Tomography, Emission-Computed, Single-Photon