Management of complications of Sideris transcatheter devices for atrial septal defect closure.
case_report · Level V
Where this comes from
- Record sourced from PubMed, PMID 8231212.
- No licence information is recorded for this record.
- Because redistribution is not established, this page shows the abstract only. Follow the links below for the full text.
Abstract
Various devices that can be inserted transvenously to close an ostium secundum atrial septal defect are undergoing clinical trials. Although these are safe and effective in most instances, they may occasionally dislodge or fail to "button" properly, causing migration and embolization. We report two cases in which the occluder and counteroccluder of the Sideris device for transvenous atrial septal defect occlusion (Custom Medical Devices, Amarillo, Tex.) failed to "button" appropriately, migrating in the right atrium in one patient and embolizing to the pulmonary artery in the second patient. An emergency operation was required to retrieve the device and repair the atrial septal defect.
Medical subject headings
- Foreign-Body Migration
- Heart Atria
- Heart Septal Defects, Atrial
- Prostheses and Implants