Comparative evaluation of clinically available inferior vena cava filters with an in vitro physiologic simulation of the vena cava.
biomechanical · Level V
Where this comes from
- Record sourced from PubMed, PMID 8234702.
- No licence information is recorded for this record.
- Because redistribution is not established, this page shows the abstract only. Follow the links below for the full text.
Abstract
A physiologic in vitro model of the human inferior vena cava (IVC) was developed to evaluate the clot-trapping efficiency of various IVC filters. The flow model closely simulates the physical parameters of the human IVC in fluid viscosity, specific gravity, temperature, and pulsatility; blood flow velocity, volume, and inherent turbulence; and vein orientation, wall compliance, and clot composition. Five filters--Greenfield (G), Bird's Nest (BN), Vena Tech (VT), Simon nitinol (SN), and titanium Greenfield (TG)--were compared by using two vena cava and blood clot sizes and horizontal and vertical orientation of the model. Each filter varied significantly in its clot-capturing efficiency, depending on vena cava size and orientation and emboli size. Overall, decreasing rank order in filter clot-trapping efficiency was SN, BN, VT, TG, and G. The in vitro model of the human IVC is an alternative to multicenter clinical trials in evaluating the relative clot-trapping efficiency of various IVC filters.
Medical subject headings
- Pulmonary Embolism
- Vena Cava Filters
- Vena Cava, Inferior