Quality audit of megavoltage radiotherapy units: intercomparison of dose at a reference point using a mailed TL-dosimetry system.

Davis, B; Faessler, P · Radiother Oncol · 1993

Where this comes from

Abstract

A dosimetry intercomparison based on mailed TL-dosimeters and the well proven IAEA/EORTC phantom is described. Its aim is to identify discrepancies in dosimetry larger than +/- 3%. Dosimeters were mailed to all radiotherapy centres in Switzerland for irradiation with 2 Gy at a reference point in a water container, using photons and electrons. Thirty-six beams were monitored. The results show an agreement of within 2% for the majority of the beams monitored. Two electron beams were at 6% of the reference value.