A finite element analysis of the effect of pin distribution on the rigidity of a unilateral external fixation system.
biomechanical · Level V
Where this comes from
- Record sourced from PubMed, PMID 8244544.
- No licence information is recorded for this record.
- Because redistribution is not established, this page shows the abstract only. Follow the links below for the full text.
Abstract
The effect of pin distribution on the rigidity of a unilateral external fixation system has been investigated using finite element techniques. The results show that spreading the pins out along the entire length of the bone is more stable than grouping the pins in clusters as is customary.
Medical subject headings
- Bone Nails
- External Fixators
- Fracture Fixation