A comparison of the Coulter STKS, Coulter S+IV, and manual analysis of white blood cell differential counts in a human immunodeficiency virus-infected population.
case_control · Level III
Where this comes from
- Record sourced from PubMed, PMID 8249907.
- No licence information is recorded for this record.
- Because redistribution is not established, this page shows the abstract only. Follow the links below for the full text.
Abstract
The white blood cell differential counts obtained by manual analysis and those obtained by automated analysis using Coulter STKS and Coulter S+IV cell counters (Coulter Electronics, Hialeah, FL) in a human immunodeficiency virus (HIV)-infected population were compared. Linear correlation analyses showed the STKS to be less accurate than the S+IV in determinations of the lymphocyte and granulocyte count. This finding was confirmed by difference analyses and by determining the number of measurements that fell outside of 95% confidence limits. Similar analyses were also done on patients not known to be HIV infected, and there was no disparity found between the STKS and S+IV in that group. In the HIV-infected population, the discrepancy between the manual and automated granulocyte counts performed on the STKS, but not on the S+IV, increased with increasing red blood cell mean corpuscular volume. Physicians caring for, and investigators performing studies on, HIV-infected patients should be aware of the potential for falsely low granulocyte numbers when using the differential from the STKS.
Medical subject headings
- HIV Infections
- Leukocyte Count