Comparison of end-tidal and arterial carbon dioxide measurements during anaesthesia with the laryngeal mask airway.
Where this comes from
- Record sourced from PubMed, PMID 8251289.
- No licence information is recorded for this record.
- Because redistribution is not established, this page shows the abstract only. Follow the links below for the full text.
Abstract
We have confirmed the value of measurement of end-tidal carbon dioxide concentration as an indicator of arterial carbon dioxide tension during the use of the laryngeal mask airway in healthy patients breathing spontaneously. The mean difference between arterial and end-tidal carbon dioxide tension was 0.52 kPa (range 0-1.5 kPa), which is similar to the difference which has been reported when a tracheal tube has been used.