Efficiency of breathing systems A and D in the Carden Ventmasta ventilator.
case_control · Level III
Where this comes from
- Record sourced from PubMed, PMID 8251292.
- No licence information is recorded for this record.
- Because redistribution is not established, this page shows the abstract only. Follow the links below for the full text.
Abstract
We have compared the efficiency of the enclosed Magill attachment (System A) and the co-axial System D (Bain) in the Carden Ventmasta ventilator using both systems, each under five different ventilatory conditions, in each of five anaesthetized patients. Efficiency was assessed in terms of the effective alveolar ventilation as a fraction of the fresh gas flow. For System A, efficiency increased from a mean of 0.37 when the total ventilation was only 50% of the fresh gas flow, to a mean of 0.74 when ventilation was 2.3 times the fresh gas flow. The efficiency was substantially and significantly less with System D: 75% of that for System A at the smaller total ventilation (95% CI 65-85%) and 65% at the larger (95% CI 59-71%). A critical examination is made of conflicting definitions and terminology of the efficiency of breathing systems.
Medical subject headings
- Anesthesia, Closed-Circuit
- Anesthesia, Inhalation
- Respiration
- Ventilators, Mechanical