Continuous spirometry: an aid to monitoring ventilation during operation.
case_report · Level V
Where this comes from
- Record sourced from PubMed, PMID 8251293.
- No licence information is recorded for this record.
- Because redistribution is not established, this page shows the abstract only. Follow the links below for the full text.
Abstract
We present six case-reports of patients who experienced inadequate ventilation as a result of endobronchial or oesophageal intubation, or obstruction to the tracheal tube or airway and were monitored with on-line spirometry. The continuously displayed pressure-volume or flow-volume loops may be compared with previously recorded baseline loops. The changing configuration of the curves offers additional and instantaneous information about the cause of increased inspiratory airway pressure, decreased compliance or increased airway resistance. Monitoring flow-volume and pressure-volume loops in conjunction with currently available techniques provides a comprehensive method of monitoring ventilation.
Medical subject headings
- Intraoperative Complications
- Respiratory Insufficiency