Conductance volumetric model of an eccentrically positioned catheter within a three-compartment ellipsoidal ventricle.
basic_science · Level V
Where this comes from
- Record sourced from PubMed, PMID 8262541.
- No licence information is recorded for this record.
- Because redistribution is not established, this page shows the abstract only. Follow the links below for the full text.
Abstract
This paper presents a mathematical model of the electric field generated and resistance measured by a conductance catheter within a three-compartment ellipsoidal approximation of a cardiac ventricle. The model is based on a novel combination of analytical and numerical techniques. The numerical component of the model predicts displacement of the catheter along the central longitudinal ventricular axis, as well as eccentric positions off that axis. The analytical component of the model enables interactive definition of model parameters during simulation. This versatile combination enables performance of validation studies that include simulations of various catheter movements within a ventricle of changing geometry. The results of our simulation studies agree with the results of previous investigators studying ex-vivo systems.
Medical subject headings
- Computer Simulation
- Models, Cardiovascular
- Models, Theoretical