Surface analysis of titanium implants.
basic_science · Level V
Where this comes from
- Record sourced from PubMed, PMID 8268381.
- No licence information is recorded for this record.
- Because redistribution is not established, this page shows the abstract only. Follow the links below for the full text.
Abstract
Medical grade titanium samples were examined using X-ray photoelectron spectroscopy before and after immersion in various proteins. Additionally, an implant removed from a patient following clinical failure was examined using scanning ion and electron microscopy. The surface of the as-received samples was found to be mainly TiO2, with contaminants of H2O/OH-, calcium and nitrogen which remained after autoclaving. The immersed proteins adhered to the titanium surface, possibly via a Ca-O link. The failed clinical sample was found to be partially fibrously encapsulated with evidence of calcification. Small amounts of TiOOH were detected at the fibrous periphery, supporting the theory of Tengvall and co-workers that in vivo titanium implants are covered in a gel of this material.
Medical subject headings
- Prostheses and Implants
- Titanium