Arterial imaging with a new forward-viewing intravascular ultrasound catheter, I. Initial studies.
basic_science · Level V
Where this comes from
- Record sourced from PubMed, PMID 8313559.
- No licence information is recorded for this record.
- Because redistribution is not established, this page shows the abstract only. Follow the links below for the full text.
Abstract
Intravascular ultrasound (IVUS) of arteries is limited by the inability of current instruments to visualize beyond the catheter tip. We have developed a prototype 4-mm-diameter forward-viewing IVUS catheter (Cardiovascular Imaging Systems, Sunnyvale, Calif) that has the ability to provide B-mode cross-sectional ultrasound data for a distance of up to 2 cm distal to the catheter tip. To study the utility of this device, a 20-MHz forward-viewing IVUS catheter was used to examine 13 arterial segments (5 human femoral arteries, 1 human carotid artery, 7 canine arteries) in vitro and 1 phantom. After imaging, all data were compared with histology (Histo). In all cases, the IVUS catheter provided forward-viewing images corresponding to the arterial geometry and demonstrated vascular landmarks and atherosclerotic lesions. There was a good correlation between Histo-determined luminal diameters (LD) and IVUS-determined diameters for a distance of 14 mm ahead of the catheter tip: IVUS LD = 1.0 Histo LD + 1.3 (r = .87). These preliminary data suggest that a forward-viewing IVUS catheter is feasible, accurate, and useful for evaluation of arterial geometry distal to the catheter tip.
Medical subject headings
- Arteries
- Catheterization
- Ultrasonography