Laboratory evaluation of the Baxter Patient-Controlled Analgesia Infusor System: a disposable patient-controlled analgesia device.
basic_science · Level V
Where this comes from
- Record sourced from PubMed, PMID 8317716.
- No licence information is recorded for this record.
- Because redistribution is not established, this page shows the abstract only. Follow the links below for the full text.
Abstract
The purpose of this study was to determine the accuracy of bolus-demand volumes obtained from the Baxter Patient-Controlled Analgesia Infusor System, a disposable patient-controlled analgesia device. An automated system was used to make demands in a pattern designed to simulate those made by a patient in the first 24 postoperative hours. Following this, a pattern of varying short interdemand intervals was used to evaluate performance of the device when demands were made near the nominal lockout interval of 6 min. Finally, a prolonged array of short interdemand intervals was used to evaluate performance as the reservoir of the devices became exhausted. Ten devices were tested. One device failed; after 70 min of testing the reservoir burst. The mean delivery of the remaining nine devices for the first 24 h of testing was 92.6% of nominal (SD = 3.9%). Evaluation near the nominal lockout interval revealed a departure of delivered volume from nominal (worst = 73.9% at an interdemand interval of 6 min). As the reservoir became exhausted, delivery increased to a maximum value, ranging from 118% to 137% of nominal, before falling to zero. In general, the devices performed consistently and the departures from nominal delivery were not considered to be of clinical significance.
Medical subject headings
- Analgesia, Patient-Controlled