Direct-current injury from external pacemaker results in tissue electrolysis.

Grossi, E A; Parish, M A; Kralik, M R; Glassman, L R; Esposito, R A; Ribakove, G H; Galloway, A C; Colvin, S B · Ann Thorac Surg · 1993

case_report · Level V

Where this comes from

Abstract

In two patients undergoing open heart operations, electrochemical burns developed at the sites of connection to an external pacing system. Investigation revealed that failure of the pacing generator caused a small, continuous, direct current to pass through the patients, resulting in electrolysis at the sites of contact with the pacing and grounding wires. This electrolytic reaction was recreated in a mock pacing system and resulted in tissue injury and disintegration of the pacing wire. Guidelines to help recognize and prevent this complication are presented.