Successful implantation of cardioverter-defibrillator systems in patients with elevated defibrillation thresholds.
retrospective_cohort · Level III
Where this comes from
- Record sourced from PubMed, PMID 8335831.
- No licence information is recorded for this record.
- Because redistribution is not established, this page shows the abstract only. Follow the links below for the full text.
Abstract
. The purpose of this study was to conduct a retrospective analysis of 16 patients with high initial defibrillation thresholds in whom a three-electrode system was used to lower defibrillation thresholds and permit implantation of a cardioverter-defibrillator system. Patients with high defibrillation thresholds (> 25 J) are uncommon but may be problematic to physicians implanting cardioverter-defibrillator systems. Most conventional systems use two defibrillating electrodes, most commonly two epicardial patches. When defibrillation thresholds remain elevated despite extensive testing of a two-electrode system, a third electrode can be incorporated and tested. However, few published data exist on the use of a three-electrode system in patients with high defibrillation thresholds. After failure to achieve satisfactory defibrillation thresholds < 25 J with a two-patch electrode system, a third electrode was incorporated and tested. In all cases, two electrodes were joined to form a common cathode or anode, while a single electrode was used as the opposite polarity electrode. Various three-electrode configurations were then tested. In all 16 patients, satisfactory defibrillation thresholds were achieved and a cardioverter-defibrillator was implanted (95% confidence interval [CI] = 0% to 21%). The mean final defibrillation threshold using the revised three-electrode system was 19.5 +/- 3.7 J (p < 0.0001). A mean of 6 +/- 3 electrode configurations/patient were tested before the final configuration was selected. A total of nine different electrode configurations were used in the 16 study patients; the most common of these incorporated left and right ventricular patches as combined cathode and a superior vena cava coil (n = 5) or right atrial patch electrode (n = 3) as single anode. Patients with high initial defibrillation thresholds can generally undergo successful cardioverter-defibrillator implantation with a three-electrode system if enough electrode configurations are tested after a third electrode is incorporated.
Medical subject headings
- Defibrillators, Implantable
- Electric Countershock
- Tachycardia, Ventricular
- Ventricular Fibrillation