Solid phantom material for the dosimetry of iodine-125 seed ophthalmic plaques.
basic_science · Level V
Where this comes from
- Record sourced from PubMed, PMID 8344860.
- No licence information is recorded for this record.
- Because redistribution is not established, this page shows the abstract only. Follow the links below for the full text.
Abstract
A tissue-equivalent solid phantom material, RE-1, closely simulating the radiological attenuation and scattering properties of the human eye for the iodine-125 photon spectrum and their Compton-scattered secondary photons, was fabricated on a polyethylene base with CaCO3 and MgO as inorganic additives. A 24 mm diameter spherical phantom was made from 1.1 mm thick sheets of RE-1, and holes were drilled in which 1 mm3 TLD cubes were placed. The radial dose function g(r), which determines the dose profile on the transverse axis, was measured in a quasi-infinite phantom of RE-1. The values obtained deviate only slightly from those for a quasi-infinite phantom made from water-equivalent material.
Medical subject headings
- Brachytherapy
- Eye Neoplasms
- Iodine Radioisotopes
- Models, Structural