The Finapres 2300e finger cuff. The influence of cuff application on the accuracy of blood pressure measurement.
case_control · Level III
Where this comes from
- Record sourced from PubMed, PMID 8346779.
- No licence information is recorded for this record.
- Because redistribution is not established, this page shows the abstract only. Follow the links below for the full text.
Abstract
Blood pressure measurements associated with malapplication of a finger cuff were compared with contemporaneous intra-arterial pressure data in seven volunteers to determine the influence of cuff application on the accuracy of the Finapres 2300e. Systolic readings in all cuff positions differed from arterial line data by more than the recommended standard and tight and loose cuff applications under and over-read respectively, in all subjects. The results show the Finapres to be sensitive to small degrees of finger cuff malapplication which contribute to the bias on direct arterial comparison and limit the reliability of the instrument in clinical practice.