Generation of x rays for medical imaging by high-power lasers: preliminary results.

Herrlin, K; Svahn, G; Olsson, C; Pettersson, H; Tillman, C; Persson, A; Wahlström, C G; Svanberg, S · Radiology · 1993

basic_science · Level V

Where this comes from

Abstract

To evaluate the use of x-ray imaging performed with a high-power laser system in biologic and medical studies. A compact terawatt laser system based on chirped pulse amplification in titanium-doped sapphire was used. X rays were generated by irradiating a high-atomic-number target (tantalum). When photons with an energy below 10 keV were removed with use of 3 mm of aluminum, the half-value layer in aluminum for the remaining x rays was approximately 10 mm. The x-ray source allowed performance of biologic magnification radiography. Star-pattern tests indicated an equivalent focal spot size of less than 60 microns. Exposures of a single pulse could be obtained. The duration of a single x-ray pulse was estimated to be of the order of picoseconds. With use of subpicosecond laser pulses, x-ray generation can occur with a smaller equivalent focal spot size than with conventional x-ray sources.