Receiver operating characteristic analysis of fracture and pneumonia detection: comparison of laser-digitized workstation images and conventional analog radiographs.
case_control · Level III
Where this comes from
- Record sourced from PubMed, PMID 8416576.
- No licence information is recorded for this record.
- Because redistribution is not established, this page shows the abstract only. Follow the links below for the full text.
Abstract
Forty pneumonia and 40 fracture cases were matched by patient age and gender with normal cases. Pediatric cases constituted 25% of this sample. All 160 examinations were laser-digitized at a spot size of 210 microns, a contrast resolution of 12 bits, and a spatial resolution of 2.35 line pairs per millimeter. The digitized images were transmitted 5 miles over a dedicated telephone line at a T1 rate from Francis Scott Key Hospital to Johns Hopkins Hospital (both in Baltimore, Md). At Johns Hopkins Hospital, eight radiologists interpreted the original analog radiograph and corresponding soft-copy display (1,280 x 1,024 pixels). Findings, confidence ratings, and image quality were reported for each reading. A receiver operating characteristic (ROC) analysis was conducted to compare observer performance under the analog and digital reading conditions. The overall sensitivity of the analog method was 89% versus 78% for the digital method (P < .001), while the specificity values were 96% and 92%, respectively (P > .20). ROC analysis similarly indicated a statistically significant difference in favor of analog radiographs (P = .028).
Medical subject headings
- Fractures, Bone
- Pneumonia
- Radiographic Image Enhancement
- Radiology Information Systems