Clinical experience with ventricular assist systems in Japan.
case_series · Level IV
Where this comes from
- Record sourced from PubMed, PMID 8417695.
- No licence information is recorded for this record.
- Because redistribution is not established, this page shows the abstract only. Follow the links below for the full text.
Abstract
In Japan, five types of ventricular assist system, including the government-approved National Cardiovascular Center type and the Tokyo University type, were applied to 202 patients as of September 30, 1991. Adult-sized ventricular assist systems were used in 194 patients and a pediatric one in 8. The major primary diagnosis was ischemic heart disease (52.5%). Sixty-three patients in cardiogenic shock with acute myocardial infarction were treated using a ventricular assist system, and the results were fairly good. The predominant indication was failure to be weaned from cardiopulmonary bypass (72.8%), and there was no case of bridging to heart transplantation. Left ventricular assist was performed in most patients (85.1%). The duration of ventricular assist system application ranged from 1 hour to 70 days, and the mean duration was 6.8 +/- 9.6 days. Duration of circulatory support had no relation to the results, and the longest support in survivors was 25 days. The weaning rate was 50.0%, and the survival rate was 26.2%. Main causes of death were heart failure, including unrecoverable heart failure, and multiple organ failure. It should be kept in mind that the decision to use a ventricular assist system should be made quickly, before major organs, including the heart itself, are irreversibly damaged.
Medical subject headings
- Heart Failure
- Heart-Assist Devices